The cornea shapes the reading:a teaching model
Fig. 1. The puff sketched like a Scheimpflug frame. Ink is deformation, red is air pressure. Simplified teaching model. Not for clinical use.
The reading
Presets are illustrative, not patient data. Thin is post-LASIK or SMILE-like. KC-like is keratoconus-like, soft and thin. Stiffness is relative, 1.00 is typical. Presets leave true IOP as set.
The air puff pushes the cornea inward. It flattens at first applanation (A1), inverts to highest concavity (HC), and recovers through second applanation (A2). The reading is taken at A1. To flatten the cornea, the air must overcome IOP and the cornea's own resistance. A thin or soft cornea gives way early and reads low. A thick or stiff one reads high. True IOP is the same in both.
m × acceleration = P(t) − Q(u) − c × velocity, where u is apex displacement, with u = 1 at flattening.
Before A1, Q = IOP + R × u. After A1, Q = IOP + R + k(u − 1).
R = 8 × stiffness × (thickness / 545)² mmHg.
The reading is the air pressure at A1, offset once so the healthy preset reads true. Every number here comes from this simplified model. It leaves out age, corneal shape and eye movement. Simplified teaching model. Not for clinical use.
Goldmann applanation tonometry (GAT) and uncorrected air-puff IOP can underestimate IOP in thin or soft corneas and overestimate it in thick or stiff ones. After LASIK or SMILE, which thin and weaken the cornea, standard readings drop although true IOP has not changed. Keratoconus corneas are softer.
Biomechanically corrected IOP (bIOP) was developed with finite element modeling and clinical validation to reduce the dependence of IOP readings on corneal thickness, stiffness and age. This tool does not compute bIOP.
Chen, Joda, Vinciguerra, Eliasy et al. J Cataract Refract Surg, 2018. doi:10.1016/j.jcrs.2018.01.023
Eliasy et al. bIOP(v2). Current Eye Research, 2023. doi:10.1080/02713683.2022.2162087
In keratoconus, the stress-strain index (SSI) fell progressively as the disease advanced and stayed largely independent of IOP and corneal thickness. This tool does not compute SSI.
© 2026 Ashkan Eliasy. Free to use. Wherever you use it, credit the source: Ashkan Eliasy, eliasy.com. Commercial use, copying or redistribution of the tool or its code requires written permission.© 2026 Ashkan Eliasy. Free to use with credit to eliasy.com. No commercial use or redistribution without written permission.